Title :
Comparison of Drive Patterns for Single Current Source EIT in Computational Phantom
Author :
Xu, Canhua ; Dong, Xiuzhen ; Shi, Xuetao ; Fu, Feng ; Shuai, Wanjun ; Liu, Ruigang ; You, Fusheng
Author_Institution :
Dept. of Biomed. Eng., Fourth Mil. Med. Univ., Xi´´an
Abstract :
Electrical impedance tomography (EIT) estimated the electrical conductivity distribution within a body. Bipolar drive pattern appeared in most EIT studies that the current was injected with a single current source and sink. In the present study, we built a computational phantom and compared the quality of dynamic EIT image in terms of data quality, resolution and anti-noise performance associated with three different bipolar drive patterns: polar drive pattern, pseudo-polar drive pattern and adjacent drive pattern. The results demonstrated that the pseudo-polar drive pattern is optimal.
Keywords :
electric impedance imaging; medical image processing; phantoms; tomography; adjacent drive pattern; antinoise performance; bipolar drive pattern; computational phantom; data quality; dynamic EIT image quality; electrical conductivity distribution; electrical impedance tomography; image resolution; pseudopolar drive pattern; single current source EIT drive pattern; Biomedical imaging; Biomedical measurements; Conductivity; Current measurement; Drives; Electrodes; Image reconstruction; Imaging phantoms; Military computing; Tomography;
Conference_Titel :
Bioinformatics and Biomedical Engineering, 2008. ICBBE 2008. The 2nd International Conference on
Conference_Location :
Shanghai
Print_ISBN :
978-1-4244-1747-6
Electronic_ISBN :
978-1-4244-1748-3
DOI :
10.1109/ICBBE.2008.704