    Rapid image acquisition in magnetic resonance imaging (MRI) offers a number of potential benefits, such as avoiding physiological effects or scanning time on patients, overcoming physical constraints inherent within the MRI scanner, or meeting timing requirements when imaging dynamic structures and processes. In this paper, the combination of compressed sensing and parallel magnetic resonance imaging techniques was proposed in order to increase the data acquisition speed of MRI. However, each coil arrangement is connected to the reception circuit via a coaxial cable which leads to several limitations. To reduce this disadvantage, we also propose a design of digital wireless transmission system based on 802.11b standard for MRI application.
  • Keywords
    biomedical MRI; body area networks; data acquisition; data compression; wireless LAN; MRI; compress sensing; data acquisition speed; digital wireless transmission; magnetic resonance imaging; Artificial neural networks; Biomedical imaging; Coils; Image resolution; Magnetic resonance imaging; Signal resolution; Transmitters;