• DocumentCode
    1671385
  • Title

    Arterial compliance and pulse pressure MR measurements: an optimized procedure combining a phase contrast multi movie and a 1D technique

  • Author

    Stevanov, M. ; Baruthio, J. ; Gounot, D.

  • Author_Institution
    Fac. de Med., Strasbourg, France
  • fYear
    1999
  • fDate
    6/21/1905 12:00:00 AM
  • Firstpage
    363
  • Lastpage
    366
  • Abstract
    An approach to estimate the compliance and pulse pressure within a vessel segment by relating the fluid flow rate and the vessel area measured by MR phase imaging has been recently proposed. The technique is based on estimates of the vascular compliance and vessel distension, which can be obtained from the correlation of the second spatial and temporal flow rate derivatives and the rate at which the flow propagates along the vessel segment. Combining the compliance with the changing in flow rate, yields the pulse pressure waveform. The technique has been improved first with an interlaced slice and frame sequence for multi movie acquisition and secondly, with a fast robust 1D method for more accurate compliance measurements. The results are given for arterial models, showing a good agreement between the experimental pressure waveforms and the reference pressure curve
  • Keywords
    biomedical MRI; blood flow measurement; blood pressure measurement; cinematography; medical image processing; arterial compliance; arterial models; experimental pressure waveforms; fast robust one dimensional method; fluid flow rate; interlaced slice/frame sequence; magnetic resonance phase imaging; multi movie acquisition; one dimensional technique; optimized procedure; phase contrast multi movie; pulse pressure magnetic resonance measurements; pulse pressure waveform; reference pressure curve; second spatial flow rate derivative; temporal flow rate derivatives; vascular compliance; vessel area; vessel distension; vessel segment; Area measurement; Fluid flow; Fluid flow measurement; Image segmentation; Magnetic resonance imaging; Motion pictures; Phase estimation; Phase measurement; Pressure measurement; Pulse measurements;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computers in Cardiology, 1999
  • Conference_Location
    Hannover
  • ISSN
    0276-6547
  • Print_ISBN
    0-7803-5614-4
  • Type

    conf

  • DOI
    10.1109/CIC.1999.825982
  • Filename
    825982