DocumentCode :
167666
Title :
Design of dental implant system by FEA
Author :
Zhihong Mao
Author_Institution :
Coll. of Mech. Eng., Guangxi Univ., Nanning, China
fYear :
2014
fDate :
8-9 May 2014
Firstpage :
675
Lastpage :
677
Abstract :
The purpose of this study is to discuss the design of abutment and fixed screw in dental implant system by FEA. Stress and deformation distributions under bite forces are discussed in detail. The results showed that improper sizes of abutment & fixed screw will greatly increase the stress and deformation of the implant system.
Keywords :
CAD; deformation; dentistry; design engineering; fasteners; finite element analysis; force; internal stresses; production engineering computing; prosthetics; FEA; abutment design; bite force; deformation distribution; dental implant system; finite element analysis; fixed screw design; stress distribution; Biomechanics; Implants; Loading; Stress; Dental implant; Design; FEA;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electronics, Computer and Applications, 2014 IEEE Workshop on
Conference_Location :
Ottawa, ON
Type :
conf
DOI :
10.1109/IWECA.2014.6845710