DocumentCode :
1690187
Title :
Filtering beat-to-beat recordings of the high resolution electrocardiogram
Author :
Kestler, H.A. ; Höher, M. ; Schwenker, F. ; Hombach, V.
Author_Institution :
Dept. of Neural Inf. Process., Ulm Univ., Germany
fYear :
1996
Firstpage :
461
Lastpage :
464
Abstract :
The aim of this study was to empirically find a high-pass filter which is suited to high resolution beat-to-beat recordings in that it should not distort the morphology of the QRS complex, while at the same time most closely resemble the magnitude response of the 4 pole Butterworth filter which is usually applied to high resolution signal averaged recordings. The Butterworth filter is applied in a bidirectional mode to the QRS midpoint. Through this, filter ringing does not obscure the endpoints of the averaged QRS, but instead the midportion is affected by artefacts. This represents no real limitation if only the duration and the terminal part of the QRS are considered for analysis. But problems arise if sequences of QRS completes are used to analyse shape changes. The filter was designed and evaluated using a least squares error criterion and three different design procedures. The result was a linear phase FIR filter of order 48 selected among 1248 filters generated.
Keywords :
Butterworth filters; FIR filters; delay circuits; electrocardiography; high-pass filters; least squares approximations; medical signal processing; signal resolution; 4 pole Butterworth filter; QRS complex morphology; QRS midpoint; artefacts; beat-to-beat recording filtering; bidirectional mode; design procedures; filter ringing; high resolution electrocardiogram; high resolution signal averaged recordings; high-pass filter; least squares error criterion; linear phase FIR filter; magnitude response; sequences; Digital filters; Finite impulse response filter; Frequency; IIR filters; Information filtering; Information filters; Nonlinear filters; Shape; Signal analysis; Signal resolution;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Computers in Cardiology, 1996
Conference_Location :
Indianapolis, IN, USA
ISSN :
0276-6547
Print_ISBN :
0-7803-3710-7
Type :
conf
DOI :
10.1109/CIC.1996.542573
Filename :
542573
Link To Document :
بازگشت