DocumentCode :
1695305
Title :
Instrumented prosthesis for knee implants monitoring
Author :
Arami, Arash ; Simoncini, Matteo ; Atasoy, Oguz ; Hasenkamp, Willyan ; Ali, Shafqat ; Bertsch, Arnaud ; Meurville, Eric ; Tanner, Steve ; Dejnabadi, Hooman ; Leclercq, Vincent ; Renaud, Philippe ; Dehollain, Catherine ; Farine, Pierre-André ; Jolles, Brig
Author_Institution :
Ecole Poly Tech. Federate de Lausanne (EPFL), Lausanne, Switzerland
fYear :
2011
Firstpage :
828
Lastpage :
835
Abstract :
In this work we present an instrumented smart knee prosthesis for in-vivo measurement of forces and kinematics. Studying the constraints, we designed minimal sensory systems to be placed in the polyethylene part of the prosthesis. The magnetic sensors and a permanent magnet are chosen and configured to measure the relative kinematics of the prosthesis. Moreover, the strain gauges were designed to measure the forces on the polyethylene part. The kinematic and kinetic measurements on a mechanical knee simulator are validated toward reference systems. The supplementary electronics, including the A/D, amplifier, rectifier and voltage doubler are designed. Consequently, by considering the necessary power budget for all the components to be performed, the optimal coils for remote powering is investigated. The system will be packaged in the polyethylene part. Therefore, by the end we will have a smart polyethylene part which can be easily modified for different types of the knee prosthesis without changing the prosthesis design.
Keywords :
biological tissues; biomechanics; biomedical electronics; biomedical materials; biomedical measurement; force measurement; intelligent sensors; magnetic sensors; patient monitoring; polymers; prosthetics; strain gauges; amplifier; arthroplasty; biomedical electronics; force measurement; instrumented prosthesis; kinematic measurement; knee implant monitoring; magnetic sensor; mechanical knee simulator; permanent magnet; polyethylene part; rectifier; remote powering; smart knee prosthesis; strain gauge; voltage doubler; Instruments; Joints; Magnetic sensors; Prosthetics; Saturation magnetization; Sensor systems; Knee arthroplasty; anisotropic magneto resistors; instrumented prosthesis; kinematics measurement; kinetics measurement; remote powering; strain gauge;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Automation Science and Engineering (CASE), 2011 IEEE Conference on
Conference_Location :
Trieste
ISSN :
2161-8070
Print_ISBN :
978-1-4577-1730-7
Electronic_ISBN :
2161-8070
Type :
conf
DOI :
10.1109/CASE.2011.6042490
Filename :
6042490
Link To Document :
بازگشت