DocumentCode :
1700923
Title :
Room temperature 2-D X-ray imaging with the Controlled-Drift Detector: first experimental results
Author :
Castoldi, Andrea ; Cattaneo, Giuseppe ; Galimberti, Antonio ; Guazzoni, Chiara ; Rehak, Pavel ; Strüder, Lothar
Author_Institution :
Dipt. di Ingegneria Nucleare, Politecnico di Milano, Milan, Italy
Volume :
1
fYear :
2001
Firstpage :
156
Abstract :
The Controlled-Drift Detector is a single-photon counting X-ray imaging silicon detector that features excellent energy and time resolution. Its distinctive feature is the readout of the signal charge packets stored in each pixel column by means of an electrostatic field in few μs. The drift time of the charge packet identifies the pixel of incidence. Several 1-D flat images of single pixel columns have been acquired to investigate the achievable position resolution and the improvement of the energy resolution at high frame rates (up to 125 kHz). At 62.5 kHz the room-temperature energy resolution at the Mn Kα line is better than 300 eV FWHM. The first 2-D X-ray images of different masks carried out at frame frequencies in the range 10-100 kHz both with radioactive and synchrotron light sources are presented.
Keywords :
X-ray detection; X-ray imaging; position sensitive particle detectors; silicon radiation detectors; Controlled-Drift Detector; Si; charge packet; energy resolution; position resolution; room temperature 2D X-ray imaging; single-photon counting X-ray imaging silicon detector; synchrotron light sources; time resolution; Computer vision; Electrostatics; Energy resolution; Image resolution; Signal resolution; Silicon; Temperature control; X-ray detection; X-ray detectors; X-ray imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2001 IEEE
ISSN :
1082-3654
Print_ISBN :
0-7803-7324-3
Type :
conf
DOI :
10.1109/NSSMIC.2001.1008431
Filename :
1008431