    Normalisation of emission and transmission data taken with a large-area positron camera, PETRRA

  • Author

    Divoli, A. ; Erlandsson, K. ; Evans, N. ; Haddock, M.C. ; Flower, MA ; Ott, R.J.

  • Author_Institution
    Inst. of Cancer Res., R. Marsden Hosp., Sutton, UK
  • Volume
    3
  • fYear
    2001
  • Firstpage
    1263
  • Abstract
    Prior to reconstruction, emission and transmission data must be normalised to account for crystal non-uniformities and for geometric effects. The method described in this work for emission scans is based on the inversion of flood images that are acquired with ether a point or a plane source. The correction is applied in the event by backprojection (BP) of the list mode data obtained with PETRRA, a novel PET scanner that comprises two large area detectors. For transmission images linear interpolation was used unsuccessfully to predict values of the missing areas of the detectors that correspond to gaps between the crystals and the crystal support frame. Iterative reconstruction methods did not solve the problem either. Simulation data showed that more than one source position could remove circular artifacts in transmission imaging.
  • Keywords
    medical image processing; positron emission tomography; PETRRA; circular artifacts; crystal nonuniformities; crystal support frame; emission data; emission imaging; event by event backprojection; flood images; geometric effects; large-area positron camera; linear interpolation; transmission data; Cameras; Crystals; Detectors; Event detection; Floods; Image reconstruction; Interpolation; Iterative methods; Positron emission tomography; Reconstruction algorithms;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2001 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-7324-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2001.1008565
  • Filename
    1008565