• DocumentCode
    1704434
  • Title

    A phantom and simulation study on body-shields to cope with radioactivity outside the field of view in 3D PET

  • Author

    Hasegawa, T. ; Murayama, H. ; Matsuura, H. ; Yamaya, T.

  • Author_Institution
    Sch. of Allied Health Sci., Kitasato Univ., Sagamihara, Japan
  • Volume
    3
  • fYear
    2001
  • Firstpage
    1267
  • Abstract
    Basic physics characteristics of body-shields to screen out radioactivity outside the field of view were analyzed by phantom experiments and Monte-Carlo simulation. The shielding effect depended primarily on the geometrical condition as in the case of end-shields. However, body-shields became more effective by increasing the shield thickness more than 10 mm as not in the case of end-shields. This characteristic difference was related with the fractions of un-scattered and shield-scattered components. A body-shield could be as effective as an end-shield if a good geometrical condition was kept with sufficient shield thickness.
  • Keywords
    Monte Carlo methods; positron emission tomography; shielding; 3D PET; Monte-Carlo simulation; body-shields; end-shields; field of view; phantom experiments; radioactivity; shield-scattered components; Analytical models; Biomedical imaging; Electromagnetic scattering; Imaging phantoms; Medical simulation; Particle scattering; Physics; Positron emission tomography; Radiation detectors; Torso;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2001 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-7324-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2001.1008566
  • Filename
    1008566