DocumentCode
1706399
Title
Preliminary results from the AROPET
Author
Jan, Meei-Ling ; Liang, Hsing-Ching ; Huang, Shin-Wen ; Tang, Jih-Shan ; Pei, Cheng-Chih ; Yeh, Ching-Kai
Author_Institution
Inst. of Nucl. Energy Res., Taiwan
Volume
3
fYear
2001
Firstpage
1607
Abstract
AROPET (Animal ROtational PET)-a rotational PET with four detectors based on BGO crystals coupled to position sensitive photomultiplier tubes has been constructed. The effective field-of-view is 50.4 mm (transaxial) by 44.8 mm (axial), which is suitable for imaging small animals. In this system, animals can be imaged with two imaging modes, planar imaging and tomographic imaging. With the planar imaging mode, the AROPET works like a gamma camera, and a 2D planar projection image is obtained. With the tomographic imaging mode, the detectors are rotated through 90°, and 3D tomographic images are obtained. The rotating diameter of AROPET is flexible and can be adjusted between 220 mm∼380 mm. The spatial resolution at the center is 2.73 mm (FWHM) while the diameter is 220 mm. The energy resolution and counting rate performance have been measured. Initial results of phantoms and animal studies obtained from AROPET are presented also.
Keywords
biological techniques; photomultipliers; position sensitive particle detectors; positron emission tomography; solid scintillation detectors; 2.73 mm; 44.8 mm; 50.4 mm; AROPET; BGO detector; counting rate; effective field-of-view; energy resolution; gamma camera; phantoms; planar imaging; position sensitive photomultiplier; small animals; spatial resolution; tomographic imaging; Animals; Cameras; Crystals; Energy measurement; Energy resolution; Optical imaging; Photomultipliers; Position sensitive particle detectors; Positron emission tomography; Spatial resolution;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2001 IEEE
ISSN
1082-3654
Print_ISBN
0-7803-7324-3
Type
conf
DOI
10.1109/NSSMIC.2001.1008648
Filename
1008648
Link To Document