• DocumentCode
    1707647
  • Title

    Quad-HIDAC PET: comparison of four image reconstruction techniques for high resolution imaging

  • Author

    Walledge, Richard J. ; Manavaki, Roido ; Reader, Andrew J. ; Jeavons, Alan P. ; Julyan, Peter J. ; Zhao, Sha ; Hastings, David L. ; Zweit, Jamal

  • Author_Institution
    Dept. of Instrum. & Anal. Sci., Univ. of Manchester Inst. of Sci. & Technol., UK
  • Volume
    3
  • fYear
    2001
  • fDate
    6/23/1905 12:00:00 AM
  • Firstpage
    1816
  • Lastpage
    1820
  • Abstract
    The quad-HIDAC PET scanner offers unparalleled high resolution PET imaging capability (<1 mm), combined with notable count rate performance (400 kcps) and sensitivity (up to 2% with scatter correction). To some extent the image resolution realized by the system is determined by the choice of image reconstruction algorithm. This work compares 4 reconstruction techniques applied to data from the quad-HIDAC scanner: 2 iterative methods (ROPLE and 3D OSEM) and 2 analytic methods (BPF and 3D RP). These four methods were chosen so as to include both a list-mode based and a projection-based method in each category. The algorithms were compared in terms of contrast, noise and resolution for a specially designed cylindrical phantom. The results indicate that the iterative methods offer improved resolution and contrast for a given noise level compared to the analytic methods. Of these iterative methods, the new list-mode technique (ROPLE) performed better than the projection-based technique
  • Keywords
    algorithm theory; image reconstruction; iterative methods; medical image processing; positron emission tomography; analytic methods; count rate performance; high resolution PET imaging capability; image reconstruction algorithm; image reconstruction techniques; image resolution; iterative methods; list-mode based method; list-mode technique; noise; noise level; projection-based method; projection-based technique; quad-HIDAC PET scanner; scatter correction; sensitivity; specially designed cylindrical phantom; Algorithm design and analysis; Band pass filters; High-resolution imaging; Image reconstruction; Image resolution; Imaging phantoms; Iterative algorithms; Iterative methods; Positron emission tomography; Scattering;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2001 IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-7324-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2001.1008695
  • Filename
    1008695