DocumentCode :
1707876
Title :
Automatic monitoring of the infusion system in a rotary heart assist device
Author :
Yu, Yih-Choung
Author_Institution :
Dept. of Electr. & Comput. Eng., Lafayette Coll., Easton, PA, USA
fYear :
2003
Firstpage :
116
Lastpage :
117
Abstract :
A fault detection algorithm was designed to monitor the infusion system of a ventricular assist device (AB-180 Circulatory Support System [AB-180 CSS], CardiacAssist, Inc.). This algorithm measures the infusion pressure, calculates the pressure derivative, and then compares to the preset thresholds to detect leaks and kinks in the infusion line, abnormal infusion rate, and infusion pump stoppage. The thresholds were identified from an experiment simulating the possible fault conditions of the infusion system. The algorithm was implemented in a data acquisition system and verified by using different pumps and controllers.
Keywords :
artificial organs; blood pressure measurement; cardiology; computerised monitoring; patient monitoring; pumps; AB-180 Circulatory Support System; abnormal infusion rate; algorithm; controllers; data acquisition system; fault conditions; fault detection algorithm; infusion line; infusion pressure; infusion pump stoppage; infusion system automatic monitoring; kink detection; leak detection; preset thresholds; pressure derivative; pumps; rotary blood pumps; rotary heart assist device; thresholds; ventricular assist device; Algorithm design and analysis; Cascading style sheets; Computerized monitoring; Control systems; Data acquisition; Fault detection; Fault diagnosis; Heart; Leak detection; Pressure measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Bioengineering Conference, 2003 IEEE 29th Annual, Proceedings of
Print_ISBN :
0-7803-7767-2
Type :
conf
DOI :
10.1109/NEBC.2003.1216019
Filename :
1216019