DocumentCode
1711901
Title
Calibration methods for electric field probes and GTEM cells
Author
Kordi, Behzad ; Lo Vetri, J. ; Bridges, Greg ; Jeffrey, Ian
Author_Institution
Dept. of Electr. & Comput. Eng., Manitoba Univ., Winnipeg, Man., Canada
Volume
3
fYear
2004
Firstpage
908
Abstract
In this paper, two techniques are presented for in-situ calibration of an electric-field probe, in our case an asymptotic conical dipole. The first technique uses a parallel-plate probe as a reference, and measurements are performed using a spectrum analyzer. In the second technique, a network analyzer is employed to measure the two-port scattering parameters of the GTEM cell (apex of the cell) and the probe, which is installed on the floor of the cell. The antenna factor is extracted from the transmission coefficient. Measurements for both techniques are carried out in a GTEM cell. The effect of a nearby EUT on the antenna factor is investigated.
Keywords
S-parameters; TEM cells; calibration; conical antennas; dipole antennas; electric field measurement; electromagnetic compatibility; probes; EMC; EUT; GTEM cells; antenna factor extraction; asymptotic conical dipole; electric field measurement; electric field probes; electromagnetic compatibility; in-situ calibration; network analyzer; parallel-plate probe; spectrum analyzer; two-port scattering parameters; Antenna measurements; Calibration; Capacitance; Dipole antennas; Electromagnetic fields; Electromagnetic measurements; Frequency; Magnetic field measurement; Probes; TEM cells;
fLanguage
English
Publisher
ieee
Conference_Titel
Electromagnetic Compatibility, 2004. EMC 2004. 2004 InternationalSymposium on
Print_ISBN
0-7803-8443-1
Type
conf
DOI
10.1109/ISEMC.2004.1349946
Filename
1349946
Link To Document