DocumentCode :
1714593
Title :
72-pixel, diamond soft X-ray camera
Author :
Prasad, R. Raghu ; Crisman, A.C. ; Qi, N. ; Gensler, S.W. ; Krishnan, Mohan
Author_Institution :
Alameda Appl. Sci. Corp., San Leandro, CA, USA
fYear :
1999
Firstpage :
259
Abstract :
Summary form only given. This paper describes a camera that can provide 200 ps snap shots of the evolution of Z-pinch radiation, in the soft X-ray and vacuum ultraviolet, with spatial resolution <0.5 mm (with 2:1 magnification). The camera uses a 12/spl times/6 array of diamond detectors with 1 mm/spl times/1 mm pixels with 85% total open area. This detector has been fabricated and the absolute sensitivity of all the pixels has been determined on the laser plasma source at Sandia National Laboratory (courtesy R.B. Spielman, SNL). The camera was subsequently used to image the X-ray and VUV emissions from Ar long implosions on Double-EAGLE, capturing 400 ps snap shots of a Z-pinch. These images show the limitations of existing instruments used to image Z-pinch emissions. They also show evidence of "spike" and "bubble" formation in the pinch.
Keywords :
X-ray imaging; Z pinch; plasma diagnostics; 18 MA; 4 MA; 8 MA; Ar long implosions; Double-EAGLE; VUV emission; X-ray emission; Z-pinch radiation evolution; bubble formation; diamond detectors; diamond soft X-ray camera; laser plasma source; soft X-ray region; spike formation; vacuum ultraviolet region; Argon; Cameras; Detectors; Instruments; Laboratories; Plasma sources; Sensor arrays; Spatial resolution; X-ray imaging; X-ray lasers;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Plasma Science, 1999. ICOPS '99. IEEE Conference Record - Abstracts. 1999 IEEE International Conference on
Conference_Location :
Monterey, CA, USA
ISSN :
0730-9244
Print_ISBN :
0-7803-5224-6
Type :
conf
DOI :
10.1109/PLASMA.1999.829591
Filename :
829591
Link To Document :