• DocumentCode
    171535
  • Title

    Double stub resonant biosensor for glucose concentrations quantification of multiple aqueous solutions

  • Author

    Chretiennot, Thomas ; Dubuc, David ; Grenier, Katia

  • Author_Institution
    LAAS, Toulouse, France
  • fYear
    2014
  • fDate
    1-6 June 2014
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    This paper presents a miniature double-stub resonant-biosensor and associated technique to quantify the glucose concentrations of two aqueous solutions localized over the sensing area of each stub. A specific procedure and the behavior of the sensor have been developed, calibrated and validated experimentally. Seven blind filling configurations have been characterized for glucose concentrations ranging from 15 to 80 mg/L and our technique predicts the glucose concentration in each channel with a root-mean-square error of 3.2 g/L. Such a level of accuracy is the tolerance range for commercial glucometer, which outlines that the microwave technique can be used for accurate glucose analysis of multiple aqueous solutions.
  • Keywords
    biochemistry; biosensors; chemical sensors; mean square error methods; patient diagnosis; sugar; associated technique; blind filling configurations; commercial glucometer; glucose analysis; glucose concentration quantification; microwave technique; miniature double-stub resonant-biosensor; multiple aqueous solutions; root-mean-square error; sensing area; sensor behavior; tolerance range; Arrays; Biosensors; Dielectric measurement; Dielectrics; Electron tubes; Immune system; Sugar; Resonators; biosensor; glucose; liquids; microstrip; multichannels;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microwave Symposium (IMS), 2014 IEEE MTT-S International
  • Conference_Location
    Tampa, FL
  • Type

    conf

  • DOI
    10.1109/MWSYM.2014.6848570
  • Filename