Ultrasonic measuring of the change in spacing between membranes in a human thorax phantom, focused to detect pleural effusions
Author
Cruz-Prieto, J. ; Bazan, I. ; Ramirez, Adrian
Author_Institution
Inst. Politec. Nac., ESIME Zacatenco, Mexico City, Mexico
fYear
2014
fDate
7-12 April 2014
Firstpage
1
Lastpage
5
Abstract
The objective of this paper is to develop ultrasound signal processing techniques suitable to measure changes in distance between two membranes in a basic human thorax phantom. The two proposed techniques are based on correlation and phase analysis, getting an accuracy of 94.93% and 92.60% respectively regarding to the measurements obtained with the vernier caliper.
Keywords
biomembranes; medical signal detection; medical signal processing; phantoms; ultrasonic measurement; human thorax phantom; membranes; phase analysis; pleural effusion detection; ultrasonic measurement; ultrasound signal processing; vernier caliper; Conferences; Couplings; IEEE catalog; Medical services; correlation; phase analysis; pleural effusion; ultrasound;