• DocumentCode
    1728214
  • Title

    Dependence of the performance of CsI-covered microstrip plate VUV photosensors on geometry: experimental results

  • Author

    Freitas, D.S.A.P. ; Veloso, J.F.C.A. ; dos Santos, J.M.F. ; Conde, C.A.N.

  • Author_Institution
    Dept. de Fisica, Coimbra Univ., Portugal
  • Volume
    2
  • fYear
    2001
  • Firstpage
    940
  • Abstract
    The behavior of a xenon-filled microstrip gas chamber (MSGC) instrumented with a microstrip plate with eight different microstrip geometries is reported. The best energy resolution achieved for 5.9-keV and for 22.1-keV X-rays was 13.6% and 7.0%, respectively. The same microstrip plate covered with a CsI film was used in place of the photomultiplier tube to detect the vacuum ultraviolet (VUV) scintillation light produced in a driftless xenon-filled gas proportional scintillation counter. The relative amplitude, energy resolution, optical positive feedback, and photoelectron extraction efficiency are discussed. The energy resolution obtained for 5.9-keV X-rays was 11.4%, which is better than that achieved with a standard proportional counter.
  • Keywords
    X-ray detection; proportional counters; ultraviolet detectors; 22.1 keV; 5.9 keV; CsI; CsI film; CsI microstrip plate; MSGC; VUV photosensors; X-rays; Xe; amplitude; energy resolution; microstrip gas chamber; optical positive feedback; photoelectron extraction efficiency; Electron tubes; Energy resolution; Geometry; Instruments; Microstrip; Optical feedback; Optical films; Photomultipliers; Solid scintillation detectors; X-rays;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2001 IEEE
  • Conference_Location
    San Diego, CA, USA
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-7324-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2001.1009710
  • Filename
    1009710