DocumentCode :
1729263
Title :
Position sensing in a cylindrical ionization detector through use of a segmented cathode
Author :
Athanasiades, Athanasios ; Lacy, Jeffrey L. ; Sun, Liang
Author_Institution :
Proportional Technol. Inc, Houston, TX, USA
Volume :
2
fYear :
2001
Firstpage :
1151
Abstract :
Position sensing in radiation detectors is essential for imaging applications and improved energy resolution. We propose an electrode configuration for cylindrical ionization chambers that accurately senses the radial and azimuthal coordinates of interaction vertices. The cathode of the proposed design is segmented into six longitudinal strip electrodes. Signals read out from each electrode are used in a computational scheme that accurately predicts the location of interacting events inside the detector, as indicated by Monte Carlo simulations. These show that the spatial resolution offered by a 24 mm diameter detector is 1 mm at 140 keV and 0.5 mm at 511 keV (assuming a root-mean-square amplifier noise of 50 electrons). At the same noise and energy levels, full-width-at-half-maximum energy resolution is 7.1% and 2.3%, respectively. These values indicate significant improvement over conventional detectors.
Keywords :
Monte Carlo methods; cathodes; ionisation chambers; position sensitive particle detectors; 140 keV; 511 keV; FWHM; Monte Carlo simulation; cylindrical ionization detector; energy resolution; position sensing; segmented cathode; spatial resolution; strip electrodes; Cathodes; Electrodes; Electrons; Energy resolution; Event detection; Image segmentation; Ionization chambers; Radiation detectors; Spatial resolution; Strips;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2001 IEEE
ISSN :
1082-3654
Print_ISBN :
0-7803-7324-3
Type :
conf
DOI :
10.1109/NSSMIC.2001.1009754
Filename :
1009754
Link To Document :
بازگشت