Title :
Design considerations for a new solid-state gamma-camera: Soltice
Author :
Gagnon, Daniel ; Zeng, Gengsheng L. ; Links, Jonathan M. ; Griesmer, Jerome J. ; Valentino, Frank C.
Author_Institution :
Nucl. Medicine Div., Philips Med. Syst., Cleveland, OH, USA
Abstract :
This paper presents an alternative to a conventional gamma camera. The proposed system is based on the combination of a new detection material - CZT - and a new acquisition geometry - 1D detector with slat collimator. The slat collimator offers a better compromise to the usual sensitivity-resolution tradeoff. Solid-state detection offers significantly improved energy resolution. The combination of the two elements, and the development of better reconstruction tools, has the potential of bringing a new level of performance to nuclear medicine imaging. The targeted system performance includes a spatial resolution of 5 mm FWHM at 10 cm from the detector surface and an energy resolution better than 5% FWHM, both at 140 keV. While the results are still preliminary and incomplete, they more than justify further investigation.
Keywords :
gamma-ray detection; position sensitive particle detectors; radioisotope imaging; semiconductor counters; CZT detector; CdZnTe; CdZnTe detector; FWHM; Soltice; energy resolution; gamma camera; nuclear medicine; slat collimator; spatial resolution; Cameras; Collimators; Detectors; Energy resolution; Geometry; Image reconstruction; Nuclear medicine; Solid state circuits; Spatial resolution; System performance;
Conference_Titel :
Nuclear Science Symposium Conference Record, 2001 IEEE
Print_ISBN :
0-7803-7324-3
DOI :
10.1109/NSSMIC.2001.1009755