• DocumentCode
    1729439
  • Title

    Analytic determination of the pinhole collimator´s point-spread function and RMS resolution with penetration

  • Author

    Metzler, S.D. ; Bowsher, J.E. ; Greer, K.L. ; Jaszczak, R.J.

  • Author_Institution
    Med. Center, Duke Univ., Durham, NC, USA
  • Volume
    2
  • fYear
    2001
  • Firstpage
    1175
  • Abstract
    Pinhole collimators are widely used to image small organs and animals. The pinhole response function (PRF) of knife-edge pinhole collimators has been estimated previously using geometric constructions without considering penetration and using "roll-off" models that employ an exponential model for the flux. An exact expression for the PRF on the imaging plane that includes the effect of aperture penetration is derived in this paper by calculating the flux for photons passing through the aperture and those passing through the attenuating material. The PRF is then used to estimate the angular-dependent root-mean-square resolution in the directions parallel and perpendicular to the tilt of the point source. The formulas are then compared with experimental data.
  • Keywords
    biological techniques; gamma-ray detection; single photon emission computed tomography; RMS resolution; SPECT; animals; aperture penetration; gamma ray imaging; organs; penetration; pinhole collimator; pinhole response function; point-spread function; Animal structures; Apertures; Breast; Gamma rays; Geometry; Image resolution; Optical collimators; Optical imaging; Single photon emission computed tomography; Solid modeling;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2001 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-7324-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2001.1009759
  • Filename
    1009759