• DocumentCode
    1732496
  • Title

    Real-time weighted integration system for rapid calculation of parametric rate-constant images in 3-dimensional positron emission tomography

  • Author

    Iida, Hiroyuki ; Miura, Shun ; Narita, Y. ; Kanno, Issei ; Satoh, T. ; Kitamura, Kokoro ; Tanaka, Kiyoshi ; Amano, M.

  • Author_Institution
    Res. Inst. for Brain & Blood Vessels, Akita
  • Volume
    3
  • fYear
    1995
  • Firstpage
    1675
  • Abstract
    A system has been developed to rapidly calculate functional images of parametric rate constants, without dynamic scan in 3D PET, and implementation of this system was tested in a clinical study following O-15 water administration. The technique is based on the weighted-integration algorithms derived for the 2- and 3-compartment models. The system consists of large-scale memory (1 GBytes) and a micro-processor that is capable of acquiring, transferring and making arbitrary arithmetic calculations during data acquisition, providing multiple time-weighted 3D-sinograms in real-time. These sinograms are reconstructed, and inserted into operational equations, thus generating parametric rate-constant images. This technique was found to reliably provide quantitative parametric rate-constant images in 3D PET, and thus would be useful for routine clinical use of 3D PET
  • Keywords
    medical image processing; positron emission tomography; 1 GByte; 2-compartment model; 3-compartment model; 3-dimensional positron emission tomography; 3D PET; H2O; O-15 water administration; arbitrary arithmetic calculations; clinical study; functional images; medical diagnostic imaging; multiple time-weighted 3D-sinograms; nuclear medicine; parametric rate-constant images calculation; real-time weighted integration system; routine clinical use; Apertures; Arithmetic; Data acquisition; Equations; Image generation; Image reconstruction; Large-scale systems; Positron emission tomography; Real time systems; Workstations;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference Record, 1995., 1995 IEEE
  • Conference_Location
    San Francisco, CA
  • Print_ISBN
    0-7803-3180-X
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1995.501908
  • Filename
    501908