  • Abstract
    Three methods for evaluating the dynamic response of force transducers against varying force are described in this paper. In all methods, the inertial force of a mass is used as the known dynamic force, and this reference force is applied to a force transducer under test. The inertial force is measured highly accurately as the as the product of the mass and the acceleration. An aerostatic linear bearing is used to obtain linear motion with sufficiently small friction acting on the mass (i.e., the moving part of the bearing). Three experimental setups were built for the dynamic calibration against an impact force, an oscillation force and a step force. Ways of integrating the three different methods are discussed.
  • Keywords
    calibration; force sensors; aerostatic linear bearing; dynamic force calibration; dynamic response; force sensor; force transducer under test; force transducers; impact force; inertial force; linear motion; mechanical quantities measurements; optical interferometer; oscillation force; pneumatic linear bearing; reference force; step force; varying force; Acceleration; Aerodynamics; Auditory system; Calibration; Force measurement; Force sensors; Motion measurement; Optical interferometry; Optical sensors; Transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 2004. IMTC 04. Proceedings of the 21st IEEE
  • ISSN
    1091-5281
  • Print_ISBN
    0-7803-8248-X
  • Type

    conf

  • DOI
    10.1109/IMTC.2004.1351062
  • Filename
    1351062