Abstract :
We report here the design, fabrication and experiments of a hollow in-plane microneedle, with which a biosensor is integrated to extract interstitial fluid from skin for glucose measurement. The design of hollow in-plane microneedles leads to a simple fabrication process. Different from common methods for immobilizing enzyme, we use capillary force to immobilize enzyme on the surface of microelectrodes. The experimental results showed a good linearity for the characteristic curve of glucose concentration, when it varies from 0 to 500mg/dl with respect to current signal value. The sensitivity of 0.4238nA was obtained.
Keywords :
biomedical measurement; biosensors; enzymes; microelectrodes; microfluidics; sugar; biosensor; enzyme immobilization; glucose analysis; glucose concentration; glucose measurement; hollow in-plane microneedle; interstitial fluid; microelectrode surface; Biochemical analysis; Biochemistry; Biosensors; Diabetes; Electrodes; Fabrication; Gold; Microelectrodes; Sampling methods; Sugar;