    Clinical evaluation of a new infrared tympanic thermometer ”TH 700” by Rowenta or “Thermo sense” by Calor

  • Author

    Brinnel, H. ; De Wasseige, P. ; Delhomme, G. ; Dittmar, A. ; Jourdain, S.

  • Author_Institution
    Hopital de l´´Arbresle, France
  • fYear
    2000
  • fDate
    2000
  • Firstpage
    255
  • Lastpage
    258
  • Abstract
    In order to evaluate the accuracy of a new infrared tympanic thermometer (ITT) TH 700 Rowenta or Thermo Sense CALOR for core temperature monitoring, rectal (Tre) and infrared tympanic temperature (Tty) have been recorded in all present inpatients of a general hospital, in 91.7% of patients, consistent results in Tty vs. Tre could be obtained (temperature range: 35.5 to 39.7°C). When both right and left ears were explored, and then the highest value of Tty (TtyMax) was retained for records, no significant difference was found between mean (±S.E.) Tre (36.83±0.05°C) and TtyMax (36.78±0.05), Detection of maximal Tty was greatly facilitated by the outward orientation of the display screen and the possibility of making several readings within a few seconds while orientating the device into the best position for alignment with the tympanic membrane. Overall correspondence between Tes and Tty as recorded by TH 700 Rowenta or Thermo Sense Color was therefore found suitable for using this ITT in everyday clinical monitoring of core temperature
  • Keywords
    biomedical equipment; biothermics; ear; patient monitoring; thermometers; 35.5 to 39.7 C; Calor; Rowenta; TH 700; Thermo sense; clinical evaluation; core temperature monitoring; display screen; everyday clinical monitoring; general hospital; infrared tympanic thermometer; left ear; medical instrumentation; rectal temperature; right ear; temperature range; Biomembranes; Displays; Ear; Hospitals; Infrared surveillance; Patient monitoring; Teleprinting; Temperature distribution; Temperature measurement; Temperature sensors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microtechnologies in Medicine and Biology, 1st Annual International, Conference On. 2000
  • Conference_Location
    Lyon
  • Print_ISBN
    0-7803-6603-4
  • Type

    conf

  • DOI
    10.1109/MMB.2000.893784
  • Filename
    893784