• DocumentCode
    1740663
  • Title

    Eye lens doses in the head pediatric CT examinations

  • Author

    Medeiros, Regina B. ; Daros, Kellen Adriana C ; Gimenez, Ricardo A. ; Lederman, Henrique M.

  • Author_Institution
    Dept. of Diagnostic Imaging, Univ. Federal de Sao Paulo, Brazil
  • Volume
    2
  • fYear
    2000
  • fDate
    2000
  • Firstpage
    1255
  • Abstract
    We investigated the dose in the eyes of 25 pediatric patients, from 0 to 3 years, submitted to CT head exam, using TLD dosimeters inside a tube set perpendicularly to the orbito-meatal baseline. The results shown that the absorbed eye doses varied from 63.0 to 94.6 mGy with an average uncertainty of 28%. These results confirm the findings of similar studies considering the differences of protocols and scanner design
  • Keywords
    computerised tomography; diagnostic radiography; dosimetry; eye; paediatrics; 0 to 3 year; 63 to 94.6 mGy; CT head exam; TLD dosimeters; X-ray dose; absorbed dose distribution; absorbed eye doses; eye lens doses; lucite phantom; orbito-meatal baseline; pediatric patients; Computed tomography; Eyes; Head; Image quality; Lenses; Pathology; Pediatrics; Protocols; Uncertainty; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2000. Proceedings of the 22nd Annual International Conference of the IEEE
  • Conference_Location
    Chicago, IL
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-6465-1
  • Type

    conf

  • DOI
    10.1109/IEMBS.2000.897961
  • Filename
    897961