Title
Internet calibration direct to national measurement standards for automatic network analysers
Author
Dudley, R.A. ; Ridler, N.M.
Author_Institution
Centre for Electromagn. & Time Metrol., Nat. Phys. Lab., Teddington, UK
Volume
1
fYear
2001
fDate
21-23 May 2001
Firstpage
255
Abstract
At the UK´s National Physical Laboratory, the Internet has been utilised to provide highly efficient and cost-effective calibration services for automatic network analysers (ANA). Remote calibration of ANAs from two of the leading instrument manufacturers has been achieved with direct access to primary national measurement standards and procedures, via the Internet for microwave frequencies in the range 45 MHz-110 GHz
Keywords
Internet; calibration; certification; computerised instrumentation; data acquisition; measurement standards; measurement uncertainty; microwave reflectometry; network analysers; peripheral interfaces; 45 MHz to 110 GHz; DAQ; Internet calibration; accreditation; automatic network analysers; certification; control firmware; cost-effective calibration services; iPIMMS technology; measurement uncertainty; national measurement standards; remote calibration; secure sockets layer; traceability; Calibration; Electric variables measurement; Electromagnetic analysis; IP networks; Instruments; Laboratories; Measurement standards; Testing; Transmission line measurements; Web and internet services;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation and Measurement Technology Conference, 2001. IMTC 2001. Proceedings of the 18th IEEE
Conference_Location
Budapest
ISSN
1091-5281
Print_ISBN
0-7803-6646-8
Type
conf
DOI
10.1109/IMTC.2001.928822
Filename
928822
Link To Document