Abstract
We performed a new series of systematic studies of gain and rate characteristics of several micropattern gaseous detectors. In contrast to earlier studies, these measurements were done at various pressures, gas mixtures, at a wide range of primary ionization and also when the whole area of the detectors was irradiated with a high intensity X-ray beam. Several new effects were discovered common to all tested detectors, and which define fundamental limits of their operation. Results of these studies allow us to suggest some concrete ways of micropattern detectors improving
Keywords
X-ray effects; drift chambers; gas mixtures; gas mixtures; high intensity X-ray beam; high rate applications; micropattern gaseous detectors; primary ionization; Avalanche breakdown; Cathodes; Charge measurement; Current measurement; Detectors; Electrodes; Electrons; Space charge; Sparks; Wire;