• DocumentCode
    1752101
  • Title

    A simple method of energy calibration for thin plastic scintillator

  • Author

    Hohara, S. ; Saiho, F. ; Tanaka, J. ; Aoki, S. ; Uozumi, Y. ; Matoba, M.

  • Author_Institution
    Graduate Sch. of Eng., Kyushu Univ., Fukuoka, Japan
  • Volume
    1
  • fYear
    2000
  • fDate
    2000
  • Abstract
    A thin plastic scintillator is used as a trigger counter of charged particle spectrometers. Since its Compton edge is not clear, it is difficult to check up the thin plastic scintillator counters with a standard gamma-ray source. In this paper a simple method is proposed to determine the Compton edge and the energy resolution of the thin plastic scintillator by fitting the simulated spectrum to the measured. The simulated spectrum was made with a Monte Carlo program considering geometries of the plastic scintillator. This technique was examined with the plastic scintillator having thickness of 2 mm to 10 mm
  • Keywords
    Monte Carlo methods; calibration; gamma-ray detection; particle spectrometers; solid scintillation detectors; Compton edge; Monte Carlo program; charged particle spectrometers; energy calibration; energy resolution; standard gamma-ray source; thin plastic scintillator; trigger counter; Calibration; Electrons; Energy measurement; Energy resolution; Plastics; Protons; Radiation detectors; Scattering; Sea measurements; Spectroscopy;