• DocumentCode
    1752147
  • Title

    Development of hydrophones for detecting high-energy reactions in water

  • Author

    Miyachi, T. ; Goto, S. ; Hasebe, N. ; Higuchi, M. ; Hinode, F. ; Ishiwata, M. ; Kikuchi, R. ; Konno, O. ; Masumura, T. ; Matsuyama, T. ; Misaki, A. ; Nakamura, I. ; Oyamada, M. ; Sato, M. ; Tazawa, Y. ; Tezuka, C.

  • Author_Institution
    Center for Nucl. Study, Tokyo Univ., Japan
  • Volume
    1
  • fYear
    2000
  • fDate
    2000
  • Firstpage
    42606
  • Abstract
    Acoustic detectors were developed using a piezo ceramic compound PZT. The shape of the PZT detector was essential to obtain a high sensitivity. A detector of a spherically shaped shell structure, whose size was 50 mm in diameter and 2 mm thick, was fabricated. Its sensitivity was calibrated to be about 40 mV/Pa at 54 kHz. Using the hydrophone, acoustic signals generated by an electron-induced cascade shower in water was detected. Experimental results were compared with simulation data and confirmed consistency
  • Keywords
    acoustic receivers; hydrophones; piezoelectric devices; PZT; PbZrO3TiO3; acoustic detectors; electron-induced cascade shower; hydrophones; water; Acoustic signal detection; Ceramics; Frequency response; Nuclear and plasma sciences; Physics; Radiation detectors; Shape; Silver; Sonar equipment; Structural shells;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2000 IEEE
  • Conference_Location
    Lyon
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-6503-8
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2000.949309
  • Filename
    949309