DocumentCode
1757911
Title
Precise Calibration Method for Triaxial Magnetometers Not Requiring Earth’s Field Compensation
Author
Zikmund, Ales ; Janosek, Michal ; Ulvr, Michal ; Kupec, Josef
Author_Institution
Dept. of Meas., Czech Tech. Univ. in Prague, Prague, Czech Republic
Volume
64
Issue
5
fYear
2015
fDate
42125
Firstpage
1250
Lastpage
1255
Abstract
A calibration procedure for calibrations of triaxial magnetometers is presented. The procedure uses a triaxial Helmholtz coil system and an Overhauser scalar magnetometer and is performed in the Earth´s field range. The triaxial coils are first calibrated with the Overhauser magnetometer, and subsequently, a triaxial magnetometer calibration is performed. As opposed to other calibration approaches, neither Earth´s field nulling system nor movements of the magnetometer are needed. A real calibration test was carried out-the extended calibration uncertainty was better than 430 ppm in sensitivity and 0.06° in orthogonality.
Keywords
calibration; magnetometers; measurement uncertainty; Overhauser scalar magnetometer; extended calibration uncertainty; triaxial Helmholtz coil system; triaxial magnetometers; Calibration; Coils; Current measurement; Earth; Magnetometers; Standards; Uncertainty; Calibration; magnetometer; precision; triaxial; triaxial.;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2015.2395531