Title :
Multi-Frequency Electrical Impedance Tomography System With Automatic Self-Calibration for Long-Term Monitoring
Author :
Hun Wi ; Sohal, Harsh ; McEwan, Alistair L. ; Eung Je Woo ; Tong In Oh
Author_Institution :
Dept. of Biomed. Eng., Kyung Hee Univ., Yongin, South Korea
Abstract :
Electrical Impedance Tomography (EIT) is a safe medical imaging technology, requiring no ionizing or heating radiation, as opposed to most other imaging modalities. This has led to a clinical interest in its use for long-term monitoring, possibly at the bedside, for ventilation monitoring, bleeding detection, gastric emptying and epilepsy foci diagnosis. These long-term applications demand auto-calibration and high stability over long time periods. To address this need we have developed a new multi-frequency EIT system called the KHU Mark2.5 with automatic self-calibration and cooperation with other devices via a timing signal for synchronization with other medical instruments. The impedance measurement module (IMM) for flexible configuration as a key component includes an independent constant current source, an independent differential voltmeter, and a current source calibrator, which allows automatic self-calibration of the current source within each IMM. We installed a resistor phantom inside the KHU Mark2.5 EIT system for intra-channel and inter-channel calibrations of all voltmeters in multiple IMMs. We show the deterioration of performance of an EIT system over time and the improvement due to automatic self-calibration. The system is able to maintain SNR of 80 dB for frequencies up to 250 kHz and below 0.5% reciprocity error over continuous operation for 24 hours. Automatic calibration at least every 3 days is shown to maintain SNR above 75 dB and reciprocity error below 0.7% over 7 days at 1 kHz. A clear degradation in performance results with increasing time between automatic calibrations allowing the tailoring of calibration to suit the performance requirements of each application.
Keywords :
biomedical equipment; calibration; electric impedance imaging; patient monitoring; phantoms; resistors; synchronisation; KHU Mark2.5 EIT system; automatic self-calibration; bleeding detection; current source calibrator; epilepsy foci diagnosis; flexible configuration; gastric emptying; imaging modalities; impedance measurement module; independent constant current source; independent differential voltmeter; interchannel calibration; intrachannel calibration; long-term monitoring; medical imaging technology; medical instruments; multifrequency electrical impedance tomography system; reciprocity error; resistor phantom; synchronization; time 24 h; timing signal; ventilation monitoring; Automatic self-calibration; electrical impedance tomography; long-term monitoring; system stability;
Journal_Title :
Biomedical Circuits and Systems, IEEE Transactions on
DOI :
10.1109/TBCAS.2013.2256785