• DocumentCode
    1765400
  • Title

    Effect of reference electrode in the nanoFET (field-effect transistor)-based biosensor experiment

  • Author

    Yeong-Tai Seo ; Kuk Jin Jang ; Min-Ho Lee ; Woo-Kyeong Seong ; Kook-Nyung Lee ; Yong-Kweon Kim

  • Author_Institution
    Sch. of Electr. Eng. & Comput. Sci., Seoul Nat. Univ., Seoul, South Korea
  • Volume
    9
  • Issue
    12
  • fYear
    2014
  • fDate
    12 2014
  • Firstpage
    874
  • Lastpage
    876
  • Abstract
    The reference electrode in an electrolyte solution affects the surface potential of the nanoFET (field-effect transistor) channel which can result in a false signal to be measured in the FET channel during bio- and chemical sensing experiments. In this reported work, it was demonstrated that different sensing results that were obtained were affected by the choice of the reference electrode through a pH and ion concentration sensing experiment. Three types of reference electrodes were compared: the SCE (saturated calomel reference electrode), the Ag/AgCl tube and the integrated Ag/AgCl electrode. The pH sensitivity equivalent to the surface potential was evaluated: 32.3, 34.6 and 16.3 mV/pH in the case of the SCE, the refillable tube type Ag/AgCl and the integrated Ag/AgCl, respectively.
  • Keywords
    biochemistry; biosensors; electrochemical electrodes; electrochemical sensors; electrolysis; field effect transistors; pH; silver; silver compounds; surface potential; Ag-AgCl; Ag-AgCl tube; chemical sensor; electrolyte solution; field-effect transistor channel; field-effect transistor-based biosensor; integrated Ag-AgCl electrode; ion concentration; nanoFET based-biosensor; nanoFET channel; pH sensitivity; reference electrode; saturated calomel reference electrode; surface potential;
  • fLanguage
    English
  • Journal_Title
    Micro & Nano Letters, IET
  • Publisher
    iet
  • ISSN
    1750-0443
  • Type

    jour

  • DOI
    10.1049/mnl.2014.0405
  • Filename
    6992359