• DocumentCode
    1765661
  • Title

    An Analytical Geometric Calibration Method for Circular Cone-Beam Geometry

  • Author

    Jingyan Xu ; Tsui, Benjamin M. W.

  • Author_Institution
    Dept. of Radiol., Johns Hopkins Univ., Baltimore, MD, USA
  • Volume
    32
  • Issue
    9
  • fYear
    2013
  • fDate
    Sept. 2013
  • Firstpage
    1731
  • Lastpage
    1744
  • Abstract
    This work is a continuation of our previous work on geometric calibration in the circular cone-beam geometry. It is well known that seven parameters completely describe such a geometry in either flat-panel X-ray computed tomography or single pinhole SPECT imaging. Previously we developed a graphical procedure to determine the detector in-plane rotation angle independently of the other six parameters. Using the discovered geometrical relationships, in this paper we determine the remaining six parameters using the cone-beam projections of a minimum of three point objects. Our method is analytical. It makes use of the parameters of the fitted ellipse from the calibration data. The parameter estimation is accurate in the noise-free case or when there is moderate projection data truncation or shorter calibration scan range ( ≤ 360°). We perform numerical evaluations to study the robustness of the proposed method under different projection noise levels and using different data acquisition ranges. Using a full 360° scan range, the estimation accuracy and precision of our method are comparable or superior to previous methods. Using a shorter acquisition range, there may be bias in the ellipse parameters obtained by simple algebraic fitting methods. This bias will propagate to the estimated geometric parameters. Such bias can be mostly eliminated by using a more sophisticated fitting algorithm. At the same noise level, the geometric parameter estimation accuracies are comparable, but the estimation precision degrades, as the acquisition range becomes shorter.
  • Keywords
    algebra; beam handling equipment; calibration; computerised tomography; data acquisition; numerical analysis; parameter estimation; single photon emission computed tomography; analytical geometric calibration method; calibration data; circular cone-beam geometry; cone-beam projection; data acquisition; detector in-plane rotation angle; ellipse parameter; estimation accuracy; fitted ellipse; flat-panel X-ray computed tomography; geometric parameter estimation; graphical procedure; noise-free case; numerical evaluation; projection data truncation; projection noise level; simple algebraic fitting method; single pinhole SPECT imaging; Calibration; Computed tomography; Detectors; Geometry; Noise; Single photon emission computed tomography; X-ray imaging; Ellipse fitting; flat-panel X-ray computed tomography (CT); geometric calibration; homogeneous representation; image reconstruction; multi-pinhole; pinhole SPECT; projective geometry; Algorithms; Calibration; Computer Simulation; Cone-Beam Computed Tomography; Image Processing, Computer-Assisted;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.2013.2266638
  • Filename
    6530715