DocumentCode :
1768601
Title :
Performance evaluation of a low power optical wireless link for biomedical data transfer
Author :
Tianyi Liu ; Bihr, Ulrich ; Anders, Jens ; Ortmanns, Maurits
Author_Institution :
Inst. of Microelectron., Univ. of Ulm, Ulm, Germany
fYear :
2014
fDate :
1-5 June 2014
Firstpage :
870
Lastpage :
873
Abstract :
We report on an experimental low power transcutaneous optical telemetric link (TOTL) for high speed biomedical data transfer. By decreasing the bandwidth to data rate ratio, a silicon photodiode with large active size can be used to significantly increase transmission efficiency as well as misalignment tolerance. Additionally, a custom-designed transimpedance amplifier (TIA) with low input referred noise and a modified on-off keying with a custom designed low power VCSEL (vertical cavity surface emitting laser) driver are used to further reduce the required power consumption. The proposed system transmits data up to 75 Mbps through 6 mm thick tissue, in the presence of a 4 mm misalignment while achieving a BER better than 10-5, with a power consumption lower than 2.8 mW.
Keywords :
biomedical telemetry; operational amplifiers; optical links; BER; TIA; TOTL; custom-designed transimpedance amplifier; experimental low power transcutaneous optical telemetric link; high speed biomedical data transfer; low power VCSEL driver; modified on-off keying; silicon photodiode; vertical cavity surface emitting laser driver; Bandwidth; Integrated optics; Mirrors; Optical modulation; Optical pulses; Optimization; Signal to noise ratio;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Circuits and Systems (ISCAS), 2014 IEEE International Symposium on
Conference_Location :
Melbourne VIC
Print_ISBN :
978-1-4799-3431-7
Type :
conf
DOI :
10.1109/ISCAS.2014.6865274
Filename :
6865274
Link To Document :
بازگشت