DocumentCode :
1779061
Title :
Measuring Harmonic Distortion Accurately Using Spectrum Analyzer
Author :
Cao Yun ; Qiu Xinyu ; Zhang Xiyang
Author_Institution :
Dept. of Meas., Beijing Aerosp. Control Center, Beijing, China
fYear :
2014
fDate :
18-20 Sept. 2014
Firstpage :
920
Lastpage :
923
Abstract :
The measurement result might incorrect if parameter setting is unsuitable when using spectrum analyzer to measure a signal´s harmonic distortion. It is especially useful to make sure the dynamic range of the spectrum analyzer. According to the definite dynamic range, the mixer input level and attenuator setting can be known. By analyzing the example of harmonic distortion measurement using spectrum analyzer, the accurate measurement method is given.
Keywords :
distortion measurement; harmonic distortion; measurement systems; spectral analysers; attenuator setting; dynamic range; harmonic distortion measurement; mixer input level; spectrum analyzer; Attenuation measurement; Attenuators; Distortion measurement; Dynamic range; Harmonic distortion; Measurement uncertainty; Mixers; attenuator setting; dynamic range; harmonic distortion; mixer level; spectrum analyzer;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Instrumentation and Measurement, Computer, Communication and Control (IMCCC), 2014 Fourth International Conference on
Conference_Location :
Harbin
Print_ISBN :
978-1-4799-6574-8
Type :
conf
DOI :
10.1109/IMCCC.2014.193
Filename :
6995163
Link To Document :