DocumentCode
1787333
Title
A Comparative Study on Interpolation Methods in Rebinning Circular Fan-Beam Data into Parallel-Beam Data
Author
Sunsik Ryu ; Jongduk Baek
Author_Institution
Inst. of Convergence Technol., Yonsei Univ., Incheon, South Korea
fYear
2014
fDate
27-29 May 2014
Firstpage
505
Lastpage
506
Abstract
Computed Tomography (CT) scanners evolved from simple parallel-beam geometry into more complex fan-beam geometry. The rebinning mechanism to convert fan-beam projections to parallel-beam projections is one of the methods simplifying the reconstruction of the CT image. Various interpolation methods result in different numerical presentations and noisy textures in the reconstructed CT images. This paper evaluates three interpolation methods, linear, frequency domain and higher order in rebinning fan-beam data into parallel-beam data in the framework of the FBP algorithm in parallel-beam CTs. The high-contrast spatial resolution, modulation transfer function (MTF), and noise are used to quantify the image quality. The noise is expressed as the noise power spectrum (NPS) and noise variance of the reconstructed image containing only Poisson distributed noise. Our results show that the higher order interpolation, cubic-spline, is the most helpful to improve the image quality in rebinning. We further demonstrate that the rebinning data gives more satisfactory results for assessment of noise variance in comparison with the fan-beam data.
Keywords
computational geometry; computerised tomography; image reconstruction; image resolution; interpolation; medical image processing; splines (mathematics); transfer functions; CT image reconstruction; MTF; NPS; Poisson distributed noise; complex fan-beam geometry; computed tomography scanners; cubic-spline; high-contrast spatial resolution; higher order interpolation; image quality; interpolation method; modulation transfer function; noise power spectrum; parallel-beam data; parallel-beam geometry; parallel-beam projections; Computed tomography; Frequency-domain analysis; Image quality; Image reconstruction; Interpolation; Noise; Splines (mathematics); MTF; image processing; interpolation; rebinning; x-ray CT;
fLanguage
English
Publisher
ieee
Conference_Titel
Computer-Based Medical Systems (CBMS), 2014 IEEE 27th International Symposium on
Conference_Location
New York, NY
Type
conf
DOI
10.1109/CBMS.2014.112
Filename
6881948
Link To Document