Robust packaging of QCR load sensor for biosignal detection
Author :
Murozaki, Yuichi ; Sakuma, Shinya ; Arai, Fumihito
Author_Institution :
Dept. of Micro-Nano Syst. Eng., Nagoya Univ., Nagoya, Japan
fYear :
2014
fDate :
10-12 Nov. 2014
Firstpage :
1
Lastpage :
2
Abstract :
We present a super wide-measurement range load sensor with high-durability for multi-biosignal sensing in daily life. We employ quartz crystal resonator (QCR) load sensor with super wide range (from 6 × 10-5 to 30 [N]) for sensing of biosignal ranges from mN order (e.g. heartbeat) to 100 N order (e.g. body motion). Furthermore, we implement an outer case to make the sensor durable against unexpected loads in practical use. The outer case increase the durability of the sensor 88 times higher than the sensor without outer case. Finally we succeeded in simultaneous sensing of heartbeat (≈ 100 [mN]) and body motion (≈ 25 [N]) with identical sensor.
Keywords :
biomechanics; biomedical measurement; cardiology; crystal resonators; electronics packaging; pressure sensors; QCR load sensor packaging; biosignal detection; body motion sensing; heartbeat sensing; high durability load sensor; multibiosignal sensing; quartz crystal resonator; superwide measurement range load sensor; Decision support systems;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Micro-NanoMechatronics and Human Science (MHS), 2014 International Symposium on
Conference_Location :
Nagoya
Print_ISBN :
978-1-4799-6678-3
Type :
conf
DOI :
10.1109/MHS.2014.7006102
Filename :
7006102
Link To Document :
بازگشت