    This paper describes a new sensor application that provides measuring the pressure in abdominal aorta aneurisms (AAA). Using a foldable prosthesis to treat AAA, leakages can occur. Therefore, we invented a batteryless telemetric pressure monitoring system for the detection of these leakages. The system consists of an implantable capsule and an external reader station. A capacitive pressure sensor and a ferrite based micro coil is integrated in the capsule. First in-vitro tests show a transmission range of up to 120 mm.
  • Keywords
    capacitive sensors; cardiology; patient treatment; pressure sensors; prosthetics; telemetry; abdominal aorta aneurisms; capacitive pressure sensor; data transfer; ferrite based microcoil; foldable prosthesis; implantable capsule; pressure sensor capsule; telemetric energy; telemetric pressure monitoring system; Abdomen; Aneurysm; Capacitive sensors; Leak detection; Monitoring; Pressure control; Pressure measurement; Prosthetics; Sensor phenomena and characterization; Telemetry;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Solid-State Sensors, Actuators and Microsystems, 2005. Digest of Technical Papers. TRANSDUCERS '05. The 13th International Conference on
  • Print_ISBN
    0-7803-8994-8
  • Type

    conf

  • DOI
    10.1109/SENSOR.2005.1497407
  • Filename
    1497407