DocumentCode :
1801480
Title :
Effect of Sampling Rate on Data Acquisition Card to Gamma ray Energy Resolution of the Digital Spectrometer
Author :
Yonggang, Huo ; Peng, Xu ; Xiaolin, Qiu ; Yuming, Di
Author_Institution :
Xi´´an Res. Inst. of Hi-tech, Xi´´an
fYear :
2007
fDate :
Aug. 16 2007-July 18 2007
Abstract :
In this paper, the effect to gamma ray resolution of the digital spectrometer is analyzed by changing the sampling rate of digitizer. The hardware consists of Cd(Zn)Te detector, PXI-5122 high-speed digitizer and computer. The software of data acquisition is programmed mainly with Laboratory Virtual Instrument Engineering Workbench (LabVIEW). The result shows the sampling rate of digitizer is lower, the resolution is worse. The resolution is close to the best resolution when the digitizer sampling rate increases.
Keywords :
analogue-digital conversion; spectrometers; virtual instrumentation; Cd(Zn)Te detector; LabVIEW; Laboratory Virtual Instrument Engineering Workbench; PXI-5122 high-speed digitizer; data acquisition card; digital spectrometer; gamma ray energy resolution; Data acquisition; Data engineering; Detectors; Energy resolution; Hardware; Instruments; Laboratories; Sampling methods; Software; Spectroscopy; LabVIEW; digitizer; resolution; sampling rate; virtual instrument;
fLanguage :
English