• DocumentCode
    1803829
  • Title

    Comparison of coronary vessel edge detection and cross line methods to quantify the degree of stenosis as assessed by digital systems

  • Author

    Stiel, GM ; Nienaber, CA ; Lattermann, A. ; Barth, K. ; Eicker, B. ; Vogel, C. ; Meinertz, T.

  • Author_Institution
    Univ. Hospital Eppendorf, Hamburg, Germany
  • fYear
    1994
  • fDate
    25-28 Sept. 1994
  • Firstpage
    473
  • Lastpage
    475
  • Abstract
    In clinical tests, the degree of 226 stenoses in 200 patients undergoing "prima vista" PTCA was evaluated digital on-site and on-line by both, automated/manual (EDa/EDm) edge detection method and crossline stenosis method (CL). The analysis was performed by use of AWOS V4.01, an angiographic workstation for digital quantitative coronary angiography. A tight correlation between EDa and EDm was observed (r=0.99, p/spl les/0.0001). In addition, excellent correlations between diameter measurements by CL and both EDa and EDm were observed (for both: r=0.97, p/spl les/.0001). It is concluded that both, automated/manual edge detection method and crossline stenosis method compare favorably for measuring the minimal lumen diameter enabling onsite as well as decision making for interventional therapy as reliable evaluation of the morphological results after such procedures in a digital catheterization laboratory.<>
  • Keywords
    angiocardiography; diagnostic radiography; diameter measurement; edge detection; medical image processing; angiographic workstation; clinical tests; coronary vessel edge detection methods; cross line methods; crossline stenosis method; decision making; digital catheterization laboratory; interventional therapy; medical diagnostic imaging; minimal lumen diameter; morphological results; stenosis degree quantification; Angiography; Automatic testing; Catheterization; Decision making; Electronic design automation and methodology; Image edge detection; Laboratories; Medical treatment; Performance analysis; Workstations;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computers in Cardiology 1994
  • Conference_Location
    Bethesda, MD, USA
  • Print_ISBN
    0-8186-6570-X
  • Type

    conf

  • DOI
    10.1109/CIC.1994.470106
  • Filename
    470106