Title :
The flexible micro resonator for the magnetic resonance catheter
Author :
DOHI, Tetsuji ; Matsumoto, Kiyoshi ; Shimoyama, Isao
Author_Institution :
Dept. of Mechano-Infomatics, Tokyo Univ., Japan
Abstract :
This paper reports on a flexible micro resonator for a magnetic resonance (MR) catheter. The micro resonator consists of a micro coil of 3.8 μH and a micro capacitor of 0.81 nF. The micro coil and the micro capacitor fabricated on a glass wafer became flexible by etching the glass substrate wholly. This micro resonator was flexible enough to roll around the catheter. We built a simple nuclear magnetic resonance (NMR) system to evaluate the flexible micro resonator as a receiver. The simple NMR system can produce the static magnetic field of 45 mT. We confirmed that the NMR signal can be obtained by the simple NMR system.
Keywords :
biomedical NMR; catheters; etching; micromechanical resonators; 0.81 nF; 45 mT; biomedical NMR; etching; flexible micro resonator; magnetic resonance catheter; micromechanical resonators; nuclear magnetic resonance system; Biomedical imaging; Capacitors; Catheters; Coils; Glass; Magnetic field measurement; Magnetic resonance; Magnetic resonance imaging; Nuclear magnetic resonance; Radio frequency;
Conference_Titel :
Solid-State Sensors, Actuators and Microsystems, 2005. Digest of Technical Papers. TRANSDUCERS '05. The 13th International Conference on
Print_ISBN :
0-7803-8994-8
DOI :
10.1109/SENSOR.2005.1497528