Abstract
Image registration is an important processing step in fluorescence microscopy, for example in tracking or super-resolution methods. Precision localization of single fluorescent molecules from a quantum limited photon detection process, subject to Gaussian readout noise, is key to the use of single molecule microscopy. It is therefore important to know the effect that registration has on the accuracy of localizing a single molecule. Here we demonstrate a suitable approach to image registration that accounts for point-wise errors in localizing the control points typically used in fluorescence microscopy. This allows expressions for the localization errors caused by the registration process to be derived, showing dependence on the number of control points and their associated photon counts.
Keywords
Gaussian noise; image registration; image resolution; measurement errors; Gaussian readout noise; associated photon counts; fluorescence microscopy image registration; localization errors; measurement errors; point-wise errors; single fluorescent molecule precision localization; single molecule microscopy; super-resolution methods; tracking methods;
fLanguage
English
Publisher
ieee
Conference_Titel
Signals, Systems and Computers (ASILOMAR), 2012 Conference Record of the Forty Sixth Asilomar Conference on
Conference_Location
Pacific Grove, CA
ISSN
1058-6393
Print_ISBN
978-1-4673-5050-1
Type
conf
DOI
10.1109/ACSSC.2012.6489300
Filename
6489300
Link To Document