Title :
Multiparametric measurement system for detection of cardiac graft rejection
Author :
Ramos-Castro, Juan ; Bragós-Bardia, Ramon ; Garcia-González, Miguel Ángel ; Salazar, Yolocuauhtli ; Fontova, Andreu ; Bayés-Genis, Antoni ; Cinca-Cuscullola, Juan ; Rosell-Ferrer, Javier
Author_Institution :
Electron. Eng. Dept., Univ. Politecnica de Catalunya, Barcelona, Spain
Abstract :
A system for detection of cardiac graft rejection based in the temperature, ECG and electrical impedance is described. The temperature is measured with two sensors to obtain the absolute temperature and the differential between two points inside the heart. A standard ECG lead and an intracavitary electrogram are also recorded. The electrical impedance of the tissue is measured with a three electrode configuration. In order to validate the system and the method several experiments are done with healthy anesthetized pigs.
Keywords :
biomedical electrodes; biomedical measurement; electric impedance measurement; electrocardiography; medical signal processing; surgery; temperature measurement; ECG; absolute temperature; anesthetized pigs; cardiac graft rejection detection; electrical impedance; heart transplantation; intracavitary electrogram; multiparametric measurement system; signal conditioning; superficial ECG; thermistor sensor; three electrode configuration; Biomedical measurements; Biopsy; Catheters; Electrocardiography; Heart; Impedance measurement; Myocardium; Temperature measurement; Temperature sensors; Thermistors;
Conference_Titel :
Instrumentation and Measurement Technology Conference, 2004. IMTC 04. Proceedings of the 21st IEEE
Print_ISBN :
0-7803-8248-X
DOI :
10.1109/IMTC.2004.1351409