DocumentCode :
1807892
Title :
Optimizing defibrillation electrodes: automating the search for better configurations
Author :
Seneta, E.B. ; Holley, L.
Author_Institution :
Centre for Biomed. Technol., Univ. of Technol., Sydney, NSW, Australia
fYear :
1994
fDate :
25-28 Sep 1994
Firstpage :
1
Lastpage :
4
Abstract :
A program has been developed that automatically optimizes internal defibrillation electrode shape, size and position with respect to the heart according to a number of useful criteria, such as electric field uniformity, current density and power density. The program design, some preliminary numerical experiments and their results are presented
Keywords :
current density; defibrillators; medical computing; optimisation; computer program design; defibrillation electrodes optimization; electric field uniformity; electrode configuration search optimization; electrode position; electrode size; internal defibrillation electrode shape; numerical experiments; power density; Computational modeling; Cooling; Cost function; Defibrillation; Electrodes; Finite element methods; Heart; Scheduling algorithm; Shape; Simulated annealing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Computers in Cardiology 1994
Conference_Location :
Bethesda, MD
Print_ISBN :
0-8186-6570-X
Type :
conf
DOI :
10.1109/CIC.1994.470264
Filename :
470264
Link To Document :
بازگشت