• DocumentCode
    1812119
  • Title

    CZT virtual Frisch-grid detector: Principles and applications

  • Author

    Cui, Y. ; Bolotnikov, A. ; Camarda, G. ; Hossain, A. ; Yang, G. ; James, R.B.

  • Author_Institution
    Brookhaven Nat. Lab., Upton, NY
  • fYear
    2009
  • fDate
    1-1 May 2009
  • Firstpage
    1
  • Lastpage
    5
  • Abstract
    Cadmium Zinc Telluride (CdZnTe or CZT) is a very attractive material for using as room-temperature semiconductor detectors, because it has a wide bandgap and a high atomic number. However, due to the material´s poor hole mobility, several special techniques were developed to ensure its suitability for radiation detection. Among them, the virtual Frisch-grid CZT detector is an attractive option, having a simple configuration, yet delivering an outstanding spectral performance. The goal of our group in Brookhaven National Laboratory (BNL) is to improve the performance of Frisch-ring CZT detectors; most recently, that effort focused on the non-contacting Frisch-ring detector, allowing us to build an inexpensive, large-volume detector array with high energy-resolution and a large effective area. In this paper, the principles of virtual Frisch-grid detectors are described, especially BNL´s innovative improvements. The potential applications of virtual Frisch-grid detectors are discussed, and as an example, a hand-held gamma-ray spectrometer using a CZT virtual Frisch-grid detector array is introduced, which is a self-contained device with a radiation detector, readout circuit, communication circuit, and high-voltage supply. It has good energy resolution of 1.4% (FWHM of 662-keV peak) with a total detection volume of ~20 cm3. Such a portable inexpensive device can be used widely in nonproliferation applications, non-destructive detection, radiation imaging, and for homeland security. Extended systems based on the same technology have potential applications in industrial- and nuclear-medical-imaging.
  • Keywords
    II-VI semiconductors; cadmium compounds; gamma-ray spectrometers; semiconductor counters; zinc compounds; CdZnTe; communication circuit; gamma-ray spectrometer; high-voltage supply; large-volume detector array; radiation detection; radiation detector; readout circuit; room-temperature semiconductor detectors; virtual Frisch-grid detector; Atomic measurements; Cadmium compounds; Gamma ray detection; Gamma ray detectors; Photonic band gap; Radiation detector circuits; Radiation detectors; Semiconductor materials; Sensor arrays; Zinc compounds; Cadmium Zinc Telluride (CdZnTe or CZT); gamma-ray detection; spectrometer;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Systems, Applications and Technology Conference, 2009. LISAT '09. IEEE Long Island
  • Conference_Location
    Farmingdale, NY
  • Print_ISBN
    978-1-4244-2347-7
  • Electronic_ISBN
    978-1-4244-2348-4
  • Type

    conf

  • DOI
    10.1109/LISAT.2009.5031559
  • Filename
    5031559