• DocumentCode
    1814055
  • Title

    Pre- and postprocessing algorithms for the correction of position dependencies of image processing parameters in ultrasonographic prostate images

  • Author

    Huynen, A.L. ; Giesen, R.J.B. ; Aarnink, R.G. ; de la Rosette, J.J.M.C.H. ; Debruyne, F.M.J. ; Wijkstra, H.

  • Author_Institution
    Dept. of Urology, Univ. Hospital Nijmegen, Netherlands
  • fYear
    1994
  • fDate
    3-6 Nov 1994
  • Firstpage
    666
  • Abstract
    Describes position dependent correction algorithms for image processing parameters in ultrasonographic images. The parameters are calculated using the co-occurrence algorithm. By the nature of ultrasound and the rotation of a single element transducer depth- and angle dependencies occur. For angle dependencies a pre-processing algorithm is described. For depth dependencies a post-calculation correction is described
  • Keywords
    biomedical ultrasonics; image classification; medical image processing; ultrasonic transducers; angle dependency; cancer detection; co-occurrence algorithm; depth dependency; image processing parameters; position dependencies; position dependent correction algorithms; post-calculation correction; postprocessing algorithms; pre-processing algorithm; preprocessing algorithms; single element transducer rotation; ultrasonographic prostate images; ultrasound; Biological materials; Biopsy; Image analysis; Image processing; Imaging phantoms; Physics; Probes; Ultrasonic imaging; Ultrasonic transducers; Visualization;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1994. Engineering Advances: New Opportunities for Biomedical Engineers. Proceedings of the 16th Annual International Conference of the IEEE
  • Conference_Location
    Baltimore, MD
  • Print_ISBN
    0-7803-2050-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1994.411852
  • Filename
    411852