DocumentCode :
1818861
Title :
A gas proportional scintillation counter coupled to a microstrip gas chamber photosensor
Author :
Veloso, J.F.C.A. ; Dos Santos, J.M.F. ; Conde, C.A.N.
Author_Institution :
Dept. de Fisica, Coimbra Univ., Portugal
Volume :
1
fYear :
1996
fDate :
2-9 Nov 1996
Firstpage :
73
Abstract :
A P10 filled microstrip gas chamber (MSGC) is used to replace the conventional photomultiplier tube as the photosensor for a gas proportional scintillation counter (GPSC). The VUV scintillation light produced in the xenon gas of the GPSC is transmitted through a high-purity quartz window to the MSGC where it is converted to photoelectrons by a CsI photocathode deposited directly onto the surface of a microstrip plate (MSP). The microstrip plate also provides the charge amplification in a gas selected for high gain and quenching. Preliminary experimental results of a study of the performance of this concept are presented
Keywords :
X-ray detection; gas scintillation detectors; photodetectors; position sensitive particle detectors; proportional counters; CsI; CsI photocathode; P10 filled microstrip gas chamber; VUV scintillation light; Xe; Xe gas; charge amplification; gas proportional scintillation counter; high-purity quartz window; microstrip gas chamber photosensor; photoelectrons; photosensor; Cathodes; Electrodes; Filling; Microstrip; Photomultipliers; Scintillation counters; Solid scintillation detectors; X-ray detection; X-ray detectors; Xenon;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE
Conference_Location :
Anaheim, CA
ISSN :
1082-3654
Print_ISBN :
0-7803-3534-1
Type :
conf
DOI :
10.1109/NSSMIC.1996.590895
Filename :
590895
Link To Document :
بازگشت