Title :
Extraction of electrical parameters of HPGe segmented detectors
Author :
Pullia, A. ; Isocrate, R. ; Venturelli, R. ; Bazzacco, D. ; Bassini, R. ; Boiano, C.
Author_Institution :
Dept. of Phys., Univ. of Milano, Italy
Abstract :
In this paper we suggest that some key physical parameters of High-Purity Germanium (HPGe) segmented detectors, including segment capacitance and leakage current, can be derived from standard noise measurements. These parameters are relevant when designing and optimizing the front-end electronics.
Keywords :
capacitance; detector circuits; gamma-ray detection; gamma-ray spectrometers; germanium radiation detectors; leakage currents; nuclear electronics; preamplifiers; semiconductor device noise; MARS detector; front-end electronics; gamma-ray spectroscopy; high-purity germanium segmented detectors; leakage current; low-noise charge preamplifier; segment capacitance; standard noise measurements; Capacitance; Electrodes; Energy resolution; Gamma ray detection; Gamma ray detectors; Germanium; Leak detection; Mars; Preamplifiers; Testing;
Conference_Titel :
Nuclear Science Symposium Conference Record, 2003 IEEE
Print_ISBN :
0-7803-8257-9
DOI :
10.1109/NSSMIC.2003.1352101