Abstract :
The detection of neutrons with thin 6Li containing films or layers within an ionisation chamber is discussed. A detection efficiency of 50% for thermal neutrons is obtainable with three 100μm thin 6Li foils. Preliminary measurements were performed with converters which contain a 1μm thin 6LiF layer. The detector efficiency is only 3%. Pulse height distributions were measured and analysed for the gases Ar, CF4 and C2F6 at pressures between 0.5 atm and 5 atm and at voltages between 50V and 500V. Diffraction peaks from polycrystalline iron and semi-crystalline pyrolytic graphite were measured in TOF mode and are presented.
Keywords :
ionisation chambers; lithium; neutron detection; 6Li converter; detection efficiency; ionisation chamber; ionisation counter; neutron detection experiments; pulse height distributions; thermal neutrons; Argon; Counting circuits; Detectors; Diffraction; Gases; Ionization chambers; Neutrons; Performance evaluation; Pulse measurements; Voltage;