• DocumentCode
    1822511
  • Title

    High spatial resolution scintillation detector based on the H8500 photomultiplier

  • Author

    Engels, R. ; Clemens, U. ; Kemmerling, G. ; Schelten, J.

  • Author_Institution
    Forschungszentrum Juelich GmbH, Germany
  • Volume
    1
  • fYear
    2003
  • fDate
    19-25 Oct. 2003
  • Firstpage
    692
  • Abstract
    The Flat-Panel photomultiplier (Hamamatsu H8500) can be utilized as a high resolution area detector for thermal neutrons and high energy gamma rays. This detector type is useful for special neutron scattering experiments and for the PET applied to small animals. In various test measurements this will be demonstrated by coupling suitable scintillators (NaI, BGO, LiI single crystals, 6Li-glass, and LiGd-Borate) to the photocathode and by feeding the 64 output signals of the photomultiplier (PM) into an active resistor network. The four output signals from the network are transferred to the pulse processing board (UniDAQ) where the event position address is calculated and where the event storage is done by memory increment.
  • Keywords
    gamma-ray detection; neutron detection; photomultipliers; scintillation counters; Flat-Panel photomultiplier; H8500 photomultiplier; PET; high energy gamma rays; high resolution area detector; high spatial resolution scintillation detector; photocathode; special neutron scattering experiments; thermal neutrons; Energy resolution; Gamma ray detection; Gamma ray detectors; Gamma rays; Neutron spin echo; Photomultipliers; Positron emission tomography; Scintillation counters; Solid scintillation detectors; Spatial resolution;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2003 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8257-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2003.1352135
  • Filename
    1352135